Maximize your lab’s productivity with the Lab2000 Glove Box, the industry’s most dependable economic solution for inert atmosphere processing. Whether you are working in battery technology, specialty chemicals, or material science, the Lab 2000 delivers a consistent≤1ppm environment through its high-efficiency closed-loop gas circulation system. This turnkey system features an automated, regenerable purification column and a user-friendly touch-screen interface, providing a premium clean-room experience at an accessible price point.

Manual:Standard1200 Double Station Split-type Ddouble-sided Glove Box

Guaranteed Safe Checkout

1200 Double Station Split-type Double-sided Glove Box
1200 Double Station Split-type Double-sided Glove Box

[wpforms id=”1190″ title=”true” description=”Request a call back”]

Enclosure material: Stainless steel 1.4301 (SUS 304), wall thickness 3 mm

Front viewing window:8 mm tempered safety glass or 10 mm polycarbonate (optional)

Glove ports:Hard-anodized aluminum alloy or POM (acetal) ports (optional),220 mm port diameter, O-ring sealed

Gloves:Butyl rubber gloves, 0.4 mm or 0.8 mm thickness (optional), 8″ diameter

Filtration:One filter on the exhaust and one on the inlet, particle rating < 0.3 um

Illumination:LED lighting integrated in the light box above the front viewing window

Ports:Multiple spare KF40 vacuum flange ports

Atmosphere Purity: H2O≤1ppm, O2 ≤1ppm 

Leakage Rate:≤0.05vol%/h

Solenoid Valves: Modular design for reduced leakage rates and simplified maintenance/replacement.

Antechamber: Large main antechamber with sliding tray; unique door design for lightweight and effortless operation.

Purification System: GP-20 inert gas purification system.

Control System: SIEMENS micro-controller (PLC).

HMI: SIEMENS touch-screen interface for intuitive access to all functional parameters.

Data Logging: Automatic recording of system operational data.

Gas Circulation: Closed-loop gas circulation; oil-free and vacuum-tight.

Pressure Control: Foot pedal integrated for hands-free pressure regulation.

Piping & Fittings: Full stainless steel gas flow lines and high-quality fittings.

Vacuum Pump: EDWARDS RV12 vacuum pump.

Filtration: HEPA (High-Efficiency Particulate Air) filters.

Automatic Pressure Regulation: Working pressure adjustable within a range of±12 mbar

Shelving: height-adjustable internal shelves

Mobility: Adjustable heavy-duty casters for easy relocation.

Lighting: Lamp cover equipped with anti-reflective (AR) coating.

  • Glove box with all-welded stainless steel design
  • All stainless steel pipework
  • Removable front window as an entry for large equipment
  • Purifier regeneration frequency once per year
  • Automatic antechamber control
  • Main antechamber plus mini antechamber
  • Vacuum pump for antechambers
  • Oxygen Analyzer and Moisture Analyzer
  • Auto purge function
  • User friendly and simple operation: Color LCD touch panel and PLC controller
  • Energy-Save mode, automatically reducing power consumption by up to 90% during idle periods
  • Automatic regenerable H2O/O2 purifier
  • Attainable purity O2<1 ppm , H2O<1 ppm (dew point also available for moisture reading)
  • Industry leading low leak rate of <0.05% Vol/hour at -10 mbar condition
  • Stainless steel encapsulated blower with frequency converter
  • Circulation capacity more than 84 m3/h (50 cfm) at ΔP = 60 mbar (60 Hz)
  • Compatible with world-wide voltage standards
  • Integrated high vacuum feedthrough
  • Conform to CE and UL
  • Three year limited warranty, and lifetime technical support
Chamber Dimensions: 1200mm(47.2″) x 1000mm(39.37″) x 900mm(35.4″) L x W x H
Front window Dimensions:  1123mm (44.2″) x 839mm (33.0″) L x H

Image Name Quantity
Oxygen analyzer 1
Dew analyzer 1
Oil mist filter 1
Bellows hose 3
Gloves 4
Edward vacuum pump 1
Clamps 7
Glove seal rings 8